Digital with tip ejector
Fully autoclavable at 121c
Detail:
| Inaccuracy (Rel. err.%) |
Imprecision (CV %) |
|
| Cat. No. |
Vol. µL |
Div. µL |
Min. vol. |
Mid. vol. |
Max. vol. |
Min. vol. |
Mid. vol. |
Max. vol. |
Tip style |
| 822.0010Y |
1 - 10 |
0.05 |
<± 2.5% |
<± 1.5% |
<± 1.0% |
< 2.5% |
< 1.6% |
< 0.7% |
200 µL |
| 822.0020 |
2 - 20 |
0.1 |
<± 2.5% |
<± 1.5% |
<± 1.0% |
< 1.7% |
< 1.1% |
< 0.5% |
200 µL |
| 822.0100 |
10 - 100 |
0.5 |
<± 1.5% |
<± 0.9% |
<± 0.8% |
< 1.0% |
< 0.6% |
< 0.2% |
200 µL |
| 822.0200 |
20 - 200 |
1 |
<± 1.5% |
<± 0.9% |
<± 0.6% |
< 0.6% |
< 0.4% |
< 0.2% |
200 µL |
| 822.1000 |
100 - 1000 |
5 |
<± 1.5% |
<± 0.6% |
<± 0.5% |
< 0.5% |
< 0.4% |
< 0.2% |
1000 µL |
|
* Measured at 0.5 µL
Performance values obtained with bi-dist. water at constant temperature
(± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655. |