Socorex Micro/Macro Pipettor, Acura


The Acura® manual family was built with the user in mind. Instruments combine high-tech materials with exceptional user friendly ergonomics. Whether in research or routine applications they bring more than superior performance and make pipetting safe and agreeable.

- Excellence in ergonomics
- Light weight and softness in use
- Unsurpassed metrologic performance and stability
- Adjustable tip ejector* fitting most tips
- Swift-set user calibration system*
- Shock, UV-light and autoclaving resistance
- CE certified IVD 98/79 EEC
- Three-year warranty



Adjustable volume micropipettes
  Inaccuracy (Rel. err.%) Imprecision (CV %)  
Cat. No. Vol. µL Div. µL Min. vol. Mid. vol. Max. vol. Min. vol. Mid. vol. Max. vol. Tip style
ESC-825.0002 0.1 - 2 0.002 <±6.0%1) <± 4.0% <± 2.0% <5.0%1) < 3.3% < 1.5% ultra 10 µL
ESC-825.0010 0.5 - 10 0.01 <±2.5%2) <± 1.8% <± 1.0% <1.8%2) < 1.2% < 0.5% ultra 10 µL
ESC-825.0010Y 1 - 10 0.01 <± 2.5% <± 1.8% <± 1.0% < 2.5% < 1.6% < 0.7% 200 µL
ESC-825.0020 2 - 20 0.02 <± 2.5% <± 1.8% <± 1.0% < 1.7% < 1.1% < 0.5% 200 µL
ESC-825.0050 5 - 50 0.1 <± 1.5% <± 1.3% <± 1.0% < 1.0% < 0.7% < 0.4% 200 µL
ESC-825.0100 10 - 100 0.1 <± 1.5% <± 1.2% <± 0.8% < 1.0% < 0.6% < 0.2% 200 µL
ESC-825.0200 20 - 200 0.2 <± 1.5% <± 1.1% <± 0.6% < 0.6% < 0.4% < 0.2% 200 µL
ESC-825.1000 100 - 1000 1 <± 1.5% <± 1.0% <± 0.5% < 0.5% < 0.4% < 0.2% 1000 µL

Measured at 1) 0.5µL, 2) 1µL
Performance values obtained with bi-dist. water at constant temperature
(± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655.

Adjustable volume macropipettes
  Inaccuracy (Rel. err.%) Imprecision (CV %)  
Cat. No. Vol. mL Div. mL Min. vol. Mid. vol. Max. vol. Min. vol. Mid. vol. Max. vol. Tip style
ESC-835.02 0.2 - 2 0.002 <± 1.5% <± 1.0% <± 0.5% < 0.5% < 0.3% < 0.2% 2 mL
ESC-835.05 0.5 - 5 0.01 <± 1.5% <± 1.1% <± 0.6% < 0.6% < 0.5% < 0.3% 5 mL
ESC-835.10 1 - 10 0.01 <± 1.5% <± 0.7% <± 0.5% < 0.5% < 0.3% < 0.2% 10 mL

 

 

Acura 825 Micropipette Features

Engineered ergonomics and design

The challenge for our engineers in combining high tech material with ergonomic design was far above average, and they succeeded excellently. Whether in research or routine applications, the result is an instrument designed and built with the users in mind.The precision pipettes offer much more than superior performance. Enjoy exceptional convenience, comfort, safety and sturdiness, enjoy your Acura® .
  Top of Page

Smart and reliable volume setting

State-of-the-art micrometric volume setting is performed smoothly and precisely by turning push button. Finely machined click-stops and free rotating cap prevent any unwanted volume alteration.
  Top of Page

Display always visible at a glance

Located on the front of the instrument, the display window allows set volume to be visible at all times during pipetting. A comforting reassurance for the user.
  Top of Page

Tip ejection - easier than ever

Most ergonomically positioned, the large surface, soft padded ejector button provides for low-pressure activation. In addition to ease of tip ejection, the shaft height adjustment system called JustipTM (4 mm span) - controlled by efficient click-stops - allows a wide selection of tips to tightly fit the nozzle.
  Top of Page

Swift-set calibration

The swift-set user calibration system - with integrated key and locking mechanism - makes tedious procedures a thing of the past. Easy and precise, it is in full compliance with international standards and recommendations.

New seal sicker as safety protection of calibration switch.

  Top of Page

Colour coding at will

All Acura® manual pipettes have colour-coded display windows and Smartie button caps suggesting a selection of suitable tips.

Are other colour combination desirable for individual, lab or department identification? Then choose among 14 colours of Smartie, or pick the Smartie mix which contains all colours.

  Top of Page

Minimum maintenance / sterilisation

Shock, heat, chemicals and UV light resistance is provided by carefully selected materials and good construction. Consequently, maintenance requirements are close to inexistant. Thorough cleaning and sterilisation can be achieved by autoclaving the instrument fully assembled (121°C / 250 °F - 20 minutes).
  Top of Page

QC certificate

Each instrument bears its own serial number and passes strict performance control attested by an individual QC certificate.

Acura 835 Macropipette Features

Engineered ergonomics and design

The challenge for our engineers in combining high tech material with ergonomic design was far above average, and they succeeded excellently. Whether in research or routine applications, the result is an instrument designed and built with the users in mind.The precision pipettes offer much more than superior performance. Enjoy exceptional convenience, comfort, safety and sturdiness, enjoy your Acura® .
  Top of Page

Smart and reliable volume setting

State-of-the-art micrometric volume setting is performed smoothly and precisely by turning push button. Finely machined click-stops and free rotating cap prevent any unwanted volume alteration.
  Top of Page

Display always visible at a glance

Located on the front of the instrument, the display window allows set volume to be visible at all times during pipetting. A comforting reassurance for the user.
  Top of Page

Tip ejection - easier than ever

Most ergonomically positioned, the large surface, soft padded ejector button provides for low-pressure activation. In addition to ease of tip ejection, the shaft height adjustment system called JustipTM (4 mm span) - controlled by efficient click-stops - allows a wide selection of tips to tightly fit the nozzle.
  Top of Page

Swift-set calibration

The swift-set user calibration system - with integrated key and locking mechanism - makes tedious procedures a thing of the past. Easy and precise, it is in full compliance with international standards and recommendations.

New seal sicker as safety protection of calibration switch.

  Top of Page

Colour coding at will

All Acura® manual pipettes have colour-coded display windows and Smartie button caps suggesting a selection of suitable tips.

Are other colour combination desirable for individual, lab or department identification? Then choose among 14 colours of Smartie, or pick the Smartie mix which contains all colours.

  Top of Page

Nozzle protection filter

The Acura® manual 835 models accommodate a cellulose nozzle filter providing added protection of both pipette and sample. The filter prevents any aerosol contamination of the pipette and contamination carry-over to the sample. It also blocs any liquid from entering in the barrel.
  Top of Page

Pasteur pipette adapter

Allows fitting of Pasteur pipettes addition to Polypropylene tips. Adapter has two O-rings guaranteeing proper holding and tightness of standard 2 mL Pasteur pipettes (Ø 7 mm). Accessory to be ordered under Cat. No. 1.835.631 (2 mL macropipette) and Cat. No. 1.835.633 (5 mL macropipette).
  Top of Page

Minimum maintenance / sterilisation

Shock, heat, chemicals and UV light resistance is provided by carefully selected materials and good construction. Consequently, maintenance requirements are close to inexistant. Thorough cleaning and sterilisation can be achieved by autoclaving the instrument fully assembled (121°C / 250 °F - 20 minutes).
  Top of Page

QC certificate

Each instrument bears its own serial number and passes strict performance control attested by an individual QC certificate.

Tell us what you think about this product, share your opinion with other people. Please make sure that your review focus on the product. Inappropriate reviews will not be posted.

Please login as customer to write a review.

Should you need more information on the product, kindly write your details to us.

Name * :
Email * :
Contact No. :
Enquiry * :
Security Code *
: Please click here to generate code.
   